RIGEL RI7002

RIGEL RI7002

The RI7002 Photometer is used in testing the integrity, checking leakages and breakages in high efficiency filter systems (HEPA & ULPA). Using the light scattering method, the Rigel RI7002 provides real-time integrity through detection of aerosol concentration upstream and downstream.

  • Settable audible and visual alarms signal when user-defined set-points are exceeded
  • Vacuum pump for ensuring a constant flow of sampled air of 1 CFM (28.3 litres/min)
  • A system of solenoid valves allows the flow to be selected between three different sources: through an absolute filter for detecting the reference zero (CLEAR), from a position entering the filter being tested (UPSTREAM), from a position downstream of the filter being tested (DOWNSTREAM)
  • Electro-optical scattering chamber
  • Luminous source for radiation of the particles inside the scattering chamber
  • Optical system for collecting the light diffused by the particles on a photomultiplier tube
  • Electronic system consisting of linear amplifier of the signal coming from the photomultiplier
  • Microprocessor
  • Touch screen display
  • Power supply unit

Specifications

Type of deviceForward Light Scattering Photometer for aerosol concentration
Field of useHEPA/ULPA filter integrity/seal tests (ISO 14644-3)
Sampling capacity28.3 L/min (nominal)
Measurement typeConcentration measurement:
– Percentage from 0,0001% a 100.0% with respect to a reference value (10-80 μg/Litre)
– Absolute with respect to a reference equal to 100 μg/Litre
Measurement Reproducibility0,5% of reading for values between 0.01% and 100%
Dynamic rangeUp to 600 μg/L
ReagentsPAO, DEHS, Ondina
Display5″ touchscreen display
Upstream stabilityNew function to control the stability of Upstream
Scanning Probe2.8″ LCD Display
Flexible probe with 360° rotation
2.5 m connection cable and tube with Photometer
Light SourceService life > 50.000 hours
Electrical supplyExternal AC/DC power supply 120W 24V
– Input: 100-240V 50-60Hz single-phase 1.8A – Output: 24Vdc 5A
Ambient conditions during operationTemperature from +1°C to +35°C
Relative humidity <95% not condensing
Ambient storage conditionsTemperature from -20°C to +55°C
Relative humidity <95% not condensing
Dimensions (L x D x H)260x370x160 mm
Weight
11,32 Kg