PMT® BioAerosol Monitoring System
- 5 LPM flow rate, 405nm laser, one particle detector and one fluorescence
detector - Measures particle size and biologic status simultaneously
- Continuous, real-time results without need for staining or reagents (up to 6
hours while on battery power) - ISO 21501-4 compliant particle counter
- 8” touchscreen usable while wearing 2 layers of latex gloves
- Can auto-trigger a sample with the Bio-Aerosol Sampler (BAS) to capture a
traditional microbial aerosol sample - Can be used to sample compressed air when connected to the Electric High
Pressure Diffuser (EHPD)
Specifications
| Size range | 0.5µm to 25µm |
| Size channels | 0.5µm,1.0µm,2.0µm,3.0µm, 5.0µm,10.0µm |
| Laser source | Long life laser |
| Size resolution | <15% @ 0.5µm (meets ISO 21501-4) |
| Count efficiency | 50%±20% for 0.5µm,100%±10% for >0.75µm (meets ISO 21501-4 and JIS B9921 ) |
| Flow rate | 2.83LPM ±3% |
| Flow rate control | Electronic, automatic closed-loop |
| Sampling time | 10 seconds-168 hours |
| Delay | 0-99 hours 59 minutes 59 seconds |
| Cycles | 1000 samples on one location |
| Interval | 5 seconds-99 hours 59 minutes 59 seconds |
| Sampling mode | Manual, auto, cumulative count Σ/ differential count ∆ or concentration |
| Zero count | <1count/5min |
| Concentration limit | 4,000,000 particles/ft³ @10% coincidence loss |
| Exhaust | Internal HEPA filter (>99.999% @ 0.3µm) |
| Display | 8.0" LCD capacitive touch screen |
| Language | English |
| Communication | RJ45, USB, SENSER-HUB, WIFI |
| Alarm | Audible built-in alarm |
| Capture of biological contamination sample | Connect the BioAerosol Sampler(BAS) via WIFI/USB to collect the biological contamination sample in real time |
| Report | Compliant with ISO/EUGMP/CHINESEGMP/Fed Std |
| Export file | PDF file or EXCEL file |
| Data storage | 119GB |
| Data security | Authority management, authority level divide into admin, operator and supervisor |
| Data reliability | Compliant with 21 CFR Part11 |
| Auto or Off-line | |
| Dimensions (HxWxD) | 10(H) x 7.87 (W) x 10.39(D) in 255(H) x 200(W) x 264(D) mm (with handle and foot mat) |
| Weight | 12.8Ibs/5.8kg;14.9Ibs/6.8kg (incl. battery) |
| Enclosure | 316L Stainless Steel and anodized aluminum |
| Power | AC 100-240V, 50 Hz/60 Hz |
| Battery | 10.8V, 9000mAhX2, rechargeable lithium battery |
| Operating conditions | Temperature: 5°C-35°C/41°F-95°F Relative humidity: 5-90%, non-condensing |
| Storage conditions | Temperature: 0°C-40°C/32°F-104°F Relative humidity: 5-90%, non-condensing |
| Calibration frequency | Once a year |
| Warranty | 12 months (calculated from the date of product activation or six months after the date of manu facture, whichever comes first). |
| Safety | EN 61010-1:2010+A1:2019, EN 61326-1:2013, EN 61326-2:2:2013, EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1, EN 62311:2008, EN 62479: 2010 |