PMT® 9510 PoE
The 9510 PoE Remote Particle Counters are advanced, high-precision instruments designed for accurate monitoring of airborne particle concentrations. They measure particle sizes from 0.5 to 25.0 μm at a stable flow rate of 1.0 CFM (28.3 LPM), providing consistent and reliable particle counting. These counters not only meet international calibration standards ISO 21501-4 and JIS B9921 but exceed them, making them a premier solution for applications demanding the highest accuracy in particle measurement where compliance to ISO 14644-2 is a requirement. The 9510 PoE units feature six user-selectable particle size channels, expandable up to 30 channels as needed. This flexibility allows for detailed particle analysis across multiple size ranges, which is particularly important in demanding environments such as cleanrooms, laboratories, or specialized industrial settings. Channel selection enables precise particle counting and classification tailored to the specific requirements of each process. Integrated critical nozzles and flow sensors maintain a stable and constant flow rate, ensuring that measurements are performed under consistent conditions for reliable and repeatable results. These technologies support precise and efficient control of air quality and particle concentration. The 9510 PoE Particle Counters are easily integrated into building automation or facility monitoring systems using the widely adopted MODBUS TCP/IP communication protocol. This interface allows seamless connectivity with existing systems and enables remote monitoring and control, enhancing operational efficiency and flexibility in managing particle concentrations across various industrial and commercial applications. With advanced technology, high measurement accuracy, and flexible configuration, the 9510 PoE Particle Counter provides an effective solution for monitoring airborne particle concentrations and is ideally suited for critical environments where precision and reliability are essential. With the 1.0 CFM flow rate this is ideal for the pharmaceutical industry or any application that requires compliance to EU GMP Annex 1 (Manufacture of Sterile Medicinal Products) and ISO 14644-2.
Specifications
| Size Range | 0.5 µm to 25 µm |
| Size Channels | Factory calibrated at 0.5, 1.0, 2.5, 5.0, 10.0 µm |
| Number of Channels | 6 channels (up to 30 optional) |
| Counting Efficiency | 50% @ 0.5µm; 100% for particles > 0.75 µm per JIS |
| Flow Rate | 1.0 CFM (28.3 LPM) |
| Concentration Limits | 1,500,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 3,000,000 particles/ft3 @ 10% coincidence (as tested and validated) |
| Light Source | Long life laser diode |
| Zero Count | < 1 count / 5 minutes (< 1 particles / 15ft3). No fault count subtraction |
| Alarms | Channel alarms on Raw counts, concentrations or mass (alarms on environmental sensors optional) |
| Calibration | NIST traceable |
| Vacuum Requirements | External vacuum > 15” (38.1cm) of Hg |
| Airflow | Internally monitored |
| Configuration/Download | USB mini-B |
| Alarm | Alarm LED ring |
| Communication Modes | MODBUS™ TCP/IP over Ethernet or WiFi (802.11 b/g) |
| Standards | ISO 21501-4 and JIS B9921 |
| Instrument Calibration | Recommended minimum once per year |
| External Surface | Stainless steel |
| Dimensions (L x W x H) | 3.59” x 1.83” x 5.01” (9.1 cm x 4.6 cm x 12.7 cm) including probes and connectors |
| Weight | 1.08 lb. (494 grams) |
| Accessories | Operating manual and Instrument Management Software (IMS) on USB flash drive, Calibration certificate |
| Optional Accessories | Power Supply, Remote Mounting Bracket, Stainless Steel Barbed Isoprobe (1.0 CFM/28.3 LPM), Filter Assembly, and Sample & Vacuum Tubing |
| Temperature | RH Probe 32-140°F (0-60°C) ±1°F (0.5°C), 0-100% ±2% |
| Buffer Memory | 65,000 sample records (rotating buffer) including particle count data and environmental data |
| Sample Time | 1 second to 99 hours |
| Power | 7 - 15 VDC (30 watts) or PoE (compliant with IEEE802.3af) |
| Operating Conditions | 41° to 104°F (5° to 40°C) / 20% to 95% non-condensing |
| Storage Conditions | 32° to 122°F (0° to 50°C) / Up to 98% non-condensing |
| Warranty | 5 year warranty | Blower warranty : 1 year |